Geology ReportsSearch

USGS · tm5B8

Colorimetric determination of nitrate plus nitrite in water by enzymatic reduction, automated discrete analyzer methods

Abstract

This report documents work at the U.S. Geological Survey (USGS) National Water Quality Laboratory (NWQL) to validate enzymatic reduction, colorimetric determinative methods for nitrate + nitrite in filtered water by automated discrete analysis. In these standard- and low-level methods (USGS I-2547-11 and I-2548-11), nitrate is reduced to nitrite with nontoxic, soluble nitrate reductase rather than toxic, granular, copperized cadmium used in the longstanding USGS automated continuous-flow analyzer methods I-2545-90 (NWQL laboratory code 1975) and I-2546-91 (NWQL laboratory code 1979). Colorimetric reagents used to determine resulting nitrite in aforementioned enzymatic- and cadmium-reduction methods are identical. The enzyme used in these discrete analyzer methods, designated AtNaR2 by its manufacturer, is produced by recombinant expression of the nitrate reductase gene from wall cress (Arabidopsis thaliana) in the yeast Pichia pastoris. Unlike other commercially available nitrate reductases we evaluated, AtNaR2 maintains high activity at 37&deg;C and is not inhibited by high-phenolic-content humic acids at reaction temperatures in the range of 20&deg;C to 37&deg;C. These previously unrecognized AtNaR2 characteristics are essential for successful performance of discrete analyzer nitrate + nitrite assays (henceforth, DA-AtNaR2) described here. Method detection levels (or limits; MDL) estimated for standard- and low-level DA-AtNaR2 nitrate + nitrite methods were 0.02 milligrams nitrogen per liter (mg-N/L) and 0.002 mg-N/L, respectively, which are comparable to 2010 NWQL long-term MDLs of the continuous-flow analyzer, cadmium-reduction methods (henceforth, CFA-CdR) they replace. Typically, reagent-water blanks for standard- and low-level DAAtNaR2 nitrate + nitrite methods are one half MDL or less. Nitrate + nitrite concentration differences for between-day replicates were 3 percent or less at or above 5 times the MDL and were as great as 35 percent near the MDL. Typically, nitrate spike recoveries from reagent water, surface water, groundwater, and high-phenolic-content, humic-acid-amended reagent water were 100&plusmn;20 percent. In addition to operational details and performance benchmarks for these new DA-AtNaR2 nitrate + nitrite assays, this report also provides results of interference studies for common inorganic and organic matrix constituents at 1, 10, and 100 times their median concentrations in surface-water and groundwater samples submitted annually to the NWQL for nitrate + nitrite analyses. Paired t-test and Wilcoxon signed-rank statistical analyses of results determined by CFA-CdR methods and DA-AtNaR2 methods indicate that nitrate concentration differences between population means or sign ranks were either statistically equivalent to zero at the 95 percent confidence level (p &ge; 0.05) or analytically equivalent to zero-that is, when p < 0.05, concentration differences between population means or medians were less than MDLs.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Charles J. Patton, Jennifer R. Kryskalla. 2011. Colorimetric determination of nitrate plus nitrite in water by enzymatic reduction, automated discrete analyzer methods. https://doi.org/10.3133/tm5b8

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related USGS reports

dMODELS, a MATLAB software package for modeling crustal deformation near volcanic centers and active faults using Global Navigation Satellite System data—User guide

dMODELS is a MATLAB software package that implements the most common analytical models used to interpret deformation measurements near faults and active volcanic centers. This manual focuses on inversion of deformation data from the Global Navigation Satellite System (GNSS). The included case studies emphasize the GNSS inversion component of the software. Source models include pressurized spherical, spheroidal, and horizontal sill (penny-crack) magma reservoirs in a homogeneous, elastic, isotropic, flat half-space. A topography correction is available for the spherical source. Dikes and faults are described following the mathematical notation for the rectangular dislocations in a homogeneous, elastic, flat half-space. Equations have been reviewed for typographical errors present in the original literature and verified against finite-element method numerical models. GNSS data from the 2006 eruption at Augustine Volcano, Alaska; the 1998–2000 unrest at Taal Volcano, Philippines; and the 2009 earthquake in L’Aquila, Italy, are used to demonstrate the application of the software package.

Techniques and Methods

Aspergillosis (Avian) case definition for wildlife

Diagnostic laboratories receive carcasses and samples for diagnostic evaluation and pathogen/toxin detection. Case definitions bring clarity and consistency to the evaluation process. Their use within and between organizations allows more uniform reporting of diseases and etiologic agents. The intent of a case definition is to provide scientifically based criteria for determining: (a) if an individual carcass has a specific disease and degree of confidence in that diagnosis and (b) if there is evidence of a pathogen or toxin in a carcass or sample (for example, swab, tissue sample, skin scraping, blood/serum sample, environmental sample, or other). This case definition is specific to aspergillosis and applies to all avian species.

Techniques and Methods

Field sampling guidelines for developing and verifying satellite remote sensing chlorophyll a concentration and fluorescence models in inland waters

Harmful algal blooms are increasing in frequency in inland waters across the United States, resulting in a need to monitor phytoplankton bloom events to track ecosystem health and productivity. Remote sensing of chlorophyll a values offers a cost-effective and powerful method for early detection and characterization of bloom events and serves as an overall indicator of water quality and trophic state, with regular, repeated sampling of landscape-wide, high spatial resolution measurements. Field measurements are necessary for developing and verifying chlorophyll a retrieval models. For model verification, chlorophyll a concentration or fluorescence and light attenuation measurements are needed; for model development, turbidity and colored dissolved organic matter concentration measurements are additionally needed; and for model development and verification, radiometric measurements, taxonomic identification of phytoplankton, inherent optical properties, and cyanotoxin concentration are further measurements that can provide context. This report outlines detailed methods and priority considerations for collecting high-quality field data in inland waters (defined as rivers, lakes, reservoirs, estuaries, streams, and wetlands). The described methods include best practices for collecting and preparing discretely collected water samples and for calibration, maintenance, and quality assurance and quality control of field sensors. Whereas the priorities will vary between applications, some general guidelines are to collect field samples (1) as close in time to a satellite overpass as possible, (2) from representative areas of the waterbody to capture the range of spatial variability, and (3) near the surface to match remote sensing reflectance data.

Techniques and Methods