Abstract
A gas chromatograph with a flame ionization detector was used in conjunction with a stripping chamber and cold trap apparatus to measure microgram-per-litre quantities of low molecular weight hydrocarbon gases in water samples. Glass bottles with ground-glass stoppers were used to store the samples. Formalin was added to stop bacterial activity and preserve samples with no measurable losses for as much as 7 days.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
D. J. Schultz, J. F. Pankow, D. Y. Tai, D. W. Stephens, R. E. Rathbun. 1976. Determination, storage, and preservation of low molecular weight hydrocarbon gases in aqueous solution. https://pubs.usgs.gov/publication/70232253