USGS · 70197758
Laboratory tests of three Z‐Land Fairfield Nodal 5‐Hz, three‐component sensors
Abstract
We conduct a number of laboratory tests at the Albuquerque Seismological Laboratory to verify the self‐noise and fidelity in which 3 three‐component Fairfield Nodal Z‐Land, Generation 2, 5‐Hz sensors are able to record seismic signals. In addition to the incoherent self‐noise of the sensors, we estimate the sensitivity of the units in digital volts/m/s, the damping, and the free period. These three parameters allow us to completely characterize the response of the instruments. We find that the responses of all components match a mean‐derived response to within 5% of amplitude and 0.03 radians in phase. This close agreement suggests that for most applications a nominal response is suitable. We also checked the timing of the units as compared to a Quanterra Q330HR and found good agreement up to 200samples/s . Finally, we compared the results of our noise tests on these sensors to a couple of nodal sensors recently deployed at the Community Wavefield Demonstration Experiment in north‐central Oklahoma and found that local site noise and not the sensor self‐noise is a fundamental limiter in the resolution of these deployed sensors at frequencies above ∼0.1Hz .
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Adam T. Ringler, Robert E. Anthony, M.S Karplus, Austin Holland, David C. Wilson. 2018-02-28. Laboratory tests of three Z‐Land Fairfield Nodal 5‐Hz, three‐component sensors. https://doi.org/10.1785/0220170236
Cite the original work for its findings. Save a collection to share your selection of sources.