Seeking excellence: An evaluation of 235 international laboratories conducting water isotope analyses by isotope-ratio and laser-absorption spectrometry

Abstract

Rationale Water stable isotope ratios ( δ 2 H and δ 18 O values) are widely used tracers in environmental studies; hence, accurate and precise assays are required for providing sound scientific information. We tested the analytical performance of 235 international laboratories conducting water isotope analyses using dual-inlet and continuous-flow isotope ratio mass spectrometers and laser spectrometers through a water isotope inter-comparison test. Methods Eight test water samples were distributed by the IAEA to international stable isotope laboratories. These consisted of a core set of five samples spanning the common δ -range of natural waters, and three optional samples (highly depleted, enriched, and saline). The fifth core sample contained unrevealed trace methanol to assess analyst vigilance to the impact of organic contamination on water isotopic measurements made by all instrument technologies. Results For the core and optional samples ~73 % of laboratories gave acceptable results within 0.2 ‰ and 1.5 ‰ of the reference values for δ 18 O and δ 2 H, respectively; ~27 % produced unacceptable results. Top performance for δ 18 O values was dominated by dual-inlet IRMS laboratories; top performance for δ 2 H values was led by laser spectrometer laboratories. Continuous-flow instruments yielded comparatively intermediate results. Trace methanol contamination of water resulted in extreme outlier δ -values for laser instruments, but also affected reactor-based continuous-flow IRMS systems; however, dual-inlet IRMS δ -values were unaffected. Conclusions Analysis of the laboratory results and their metadata suggested inaccurate or imprecise performance stemmed mainly from skill- and knowledge-based errors including: calculation mistakes, inappropriate or compromised laboratory calibration standards, poorly performing instrumentation, lack of vigilance to contamination, or inattention to unreasonable isotopic outcomes. To counteract common errors, we recommend that laboratories include 1–2 'known' control standards in all autoruns; laser laboratories should screen each autorun for spectral contamination; and all laboratories should evaluate whether derived d- excess values are realistic when both isotope ratios are measured. Combined, these data evaluation strategies should immediately inform the laboratory about fundamental mistakes or compromised samples.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Leonard I. Wassenaar, S. Terzer-Wassmuth, C. Douence, L. Araguas-Araguas, P. K. Aggarwal, Tyler B. Coplen. 2018-02-08. Seeking excellence: An evaluation of 235 international laboratories conducting water isotope analyses by isotope-ratio and laser-absorption spectrometry. https://doi.org/10.1002/rcm.8052

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related USGS reports

Reaffirming the consistency of the original stock of NBS 19 limestone and its byproduct with finer grain size for δ13C and δ18O calibrations—Both distributed internationally

Rationale Approximately 3–4 kg of NBS 19 limestone, prepared in 1982 by the U.S. Geological Survey (USGS), serves as a primary reference for the carbon isotope scales VPDB and VPDB-LSVEC and the oxygen isotope scale VPDB. Its consensus values are +1.95‰ for δ 13 C and −2.2‰ for δ 18 O measurements. After more than three decades of distribution, the number of units labeled “NBS 19” shrunk substantially, and the remaining material was quarantined, with a small fraction retained at the USGS and a larger fraction entrusted to the International Atomic Energy Agency (IAEA) for secure storage. Given the critical role of NBS 19, verifying the integrity of its δ 13 C and δ 18 O values across these storage locations is essential. Method Samples from IAEA and USGS stock materials were selected to ensure traceability of carbon and oxygen isotopes to NBS 19 distribution units from various storage locations. To achieve this, distribution units labeled as NBS 19 from both the IAEA and the National Institute of Standards and Technology (NIST) were included. During evaluation, two distinct grain sizes were identified within these units. Grain size was assessed using sieves on five selected samples. The δ 13 C and δ 18 O values of finer and coarser fractions (both labeled NBS 19) were measured at USGS using continuous-flow isotope-ratio mass spectrometry on 100- and 200-μg samples. Results Expanded 95% measurement uncertainties ( k = 2) for δ 13 C and δ 18 O determinations are 0.027‰ and 0.038‰ or better, respectively, for quarantined NBS 19 stocks and finer byproduct calcite, which we term “NBS-19 byproduct,” regardless of storage location. Between the 1990s and circa 2011, NBS-19 byproduct labeled “NBS 19,” distributed by NIST and the IAEA, shows these findings. Conclusions NBS 19 primary isotopic reference material should be used for calibration of new isotopic reference materials rather than NBS-19 byproduct secondary isotopic reference material.

Rapid Communications in Mass Spectrometry

Three new calcium formate reference materials for δ13C measurements and a redetermination of the R(13C/12C) ratio for VPDB based on proton nuclear magnetic resonance measurements

Rationale Isotope ratio mass spectrometry (IRMS) and proton nuclear magnetic resonance ( 1 H NMR) spectroscopy are independent techniques for determining the isotope ratio R ( 13 C/ 12 C) in organic compounds. However, the lack of suitable reference materials has limited intercalibration of results from these methods. Methods Three high-purity calcium formate isotopic reference materials were developed, each ideal for accurate isotope ratio measurements by both 1 H NMR and isotope-ratio mass spectrometry (IRMS). For each material, R ( 13 C/ 12 C) was determined by 1 H NMR, and the relative isotope abundance ( δ 13 C VPDB-LSVEC ) by IRMS. The combined IRMS and NMR results were used to determine the R ( 13 C/ 12 C) value of Vienna Peedee belemnite (VPDB). Results The combined datasets yield a value for R ( 13 C/ 12 C) of 0.0111050 ± 0.0000047 ( k = 2) for VPDB, which is 6.7‰ lower than the commonly cited value of 0.011180, but similar to other recent determinations. These three calcium formate reference materials exhibit high accuracy across both analytical methods and provide robust tools for instrument calibration, method validation, and inter-laboratory comparison in carbon stable isotope ratio analysis. Conclusions An accurate R ( 13 C/ 12 C) value for VPDB strengthens the link between relative ( δ 13 C VPDB-LSVEC ) values from IRMS and R ( 13 C/ 12 C) isotope ratios from NMR and other techniques. These three calcium formate isotopic reference materials (USGS106, USGS107, and USGS108) are available from the US Geological Survey and provide reference materials for calibration, method validation, and inter-laboratory comparison in carbon isotope ratio analysis.

Rapid Communications in Mass Spectrometry

Correction to “Stable isotope reference materials and scale definitions – outcomes of the 2024 IAEA experts meeting”

This article corrects the following: Camin, F., Besic, D., Brewer, P., Allison, C., Coplen, T., Dunn, P., Gehre, M., Gröning, M., Meijer, H.-A.-J, Hélie, J.-F., Iacumin, P., Kraft, R., Krajnc, B., Kümmel, S., Lee, S., Meija, J., Mester, Z., Mohn, J., Moossen, H., Qi, H., Skrzypek, G., Sperlich, P., Viallon, J., Wassenaar, L. and Wielgosz, R. (2025). Stable Isotope Reference Materials and Scale Definitions—Outcomes of the 2024 IAEA Experts Meeting. Rapid Commun Mass Spectrom, 39: e10018, https://doi.org/10.1002/rcm.10018. In the above article, the text of note 3 in Section 4.2 is incorrect. The correct text is shown below: 3. Whilst limited amounts of the original VSMOW and SLAP scale-defining reference materials may still be available, to enable continuous wide distribution of RMs for VSMOW-SLAP scale realization (normalization, calibration) the scale-maintaining reference materials VSMOW2 and SLAP2 are available. VSMOW2 and SLAP2 have δ(18O) values identical to VSMOW and SLAP, respectively, but with an uncertainty with respect to VSMOW and SLAP. We apologize for this error. The CIAAW-IUPAC endorsed all isotope delta scale definitions appearing in Camin et al. (2025) and agreed to publish the current and all future updates and corrections to these definitions on its website (www.ciaaw.org).

Rapid Communications in Mass Spectrometry