On the absolute calibration of SO 2 cameras

Abstract

Sulphur dioxide emission rate measurements are an important tool for volcanic monitoring and eruption risk assessment. The SO 2 camera technique remotely measures volcanic emissions by analysing the ultraviolet absorption of SO 2 in a narrow spectral window between 300 and 320 nm using solar radiation scattered in the atmosphere. The SO 2 absorption is selectively detected by mounting band-pass interference filters in front of a two-dimensional, UV-sensitive CCD detector. One important step for correct SO 2 emission rate measurements that can be compared with other measurement techniques is a correct calibration. This requires conversion from the measured optical density to the desired SO 2 column density (CD). The conversion factor is most commonly determined by inserting quartz cells (cuvettes) with known amounts of SO 2 into the light path. Another calibration method uses an additional narrow field-of-view Differential Optical Absorption Spectroscopy system (NFOVDOAS), which measures the column density simultaneously in a small area of the camera’s field-of-view. This procedure combines the very good spatial and temporal resolution of the SO 2 camera technique with the more accurate column densities obtainable from DOAS measurements. This work investigates the uncertainty of results gained through the two commonly used, but quite different, calibration methods (DOAS and calibration cells). Measurements with three different instruments, an SO 2 camera, a NFOVDOAS system and an Imaging DOAS (I-DOAS), are presented. We compare the calibration-cell approach with the calibration from the NFOV-DOAS system. The respective results are compared with measurements from an I-DOAS to verify the calibration curve over the spatial extent of the image. The results show that calibration cells, while working fine in some cases, can lead to an overestimation of the SO 2 CD by up to 60% compared with CDs from the DOAS measurements. Besides these errors of calibration, radiative transfer effects (e.g. light dilution, multiple scattering) can significantly influence the results of both instrument types. The measurements presented in this work were taken at Popocatepetl, Mexico, between 1 March 2011 and 4 March 2011. Average SO 2 emission rates between 4.00 and 14.34 kg s −1 were observed.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Peter Lubcke, Nicole Bobrowski, Sebastian Illing, Christoph Kern, Jose Manuel Alvarez Nieves, Leif Vogel, Johannes Zielcke, Hugo Delgados Granados, Ulrich Platt. 2013-03-14. On the absolute calibration of SO 2 cameras. https://doi.org/10.5194/amt-6-677-2013

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related USGS reports

Evaluating mass flow meter measurements from chambers for greenhouse gas emissions from orphan wells and other point sources

This study evaluates the performance of a rigid gas flux chamber equipped with a mass flow meter (MFM) for measuring gas emissions from leaking orphan wells and similar pressure-driven gas point sources. We conducted a series of laboratory and field experiments to evaluate the sensitivity, stability, and dynamic range of an MFM chamber system and found an optimal method for sealing the chamber to the ground to isolate the emission source. From these results, we estimate the effects of different soil gas permeabilities on measurements and identify the uncertainty of environmental processes that can impact measurements. Simulations of an MFM chamber are compared to those of a dynamic flux chamber to contrast the data derived with both methodologies and illustrate the potential for measuring high variability leaks with the MFM chamber. Using a low flow resistance MFM and a chamber well-sealed to the ground, it is possible to measure leaks down to 1.08 x 10 -3 cubic meters per hour (m 3 h −1 ) (refenced to 25°/1 atm), corresponding to 0.77 grams per hour (g h −1 ) methane or 2.11 g h −1 carbon dioxide, with a mean uncertainty of 0.89 % relative standard deviation. Environmental processes such as heated gas inside the chamber from solar gain, wind blowing across the chamber vent, and changing humidity in the chamber, can cause variation in MFM measurements. Over 11 d of continuous monitoring under varying weather conditions, the standard deviation of the environmentally sourced signals was found to be 7.40 x 10 -3 m 3 h −1 (equivalent to or 5.27 g h −1 methane or 14.45 g h −1 carbon dioxide). Strategies to obtain the highest quality data from MFM chambers include burying the edges of the chamber below the surface sufficiently deep to seal the chamber edges against gas flow and soaking the dirt with water to lower the chances of escaping gases, while monitoring the gas flow and adjusting the chamber seal to achieve a maximum flow rate.

Atmospheric Measurement Techniques

Radiometric calibration of a non-imaging airborne spectrometer to measure the Greenland ice sheet surface

Methods to radiometrically calibrate a non-imaging airborne visible-to-shortwave infrared (VSWIR) spectrometer to measure the Greenland ice sheet surface are presented. Airborne VSWIR measurement performance for bright Greenland ice and dark bare rock/soil targets is compared against the MODerate resolution atmospheric TRANsmission (MODTRAN ® ) radiative transfer code (version 6.0), and a coincident Landsat 8 Operational Land Imager (OLI) acquisition on 29 July 2015 during an in-flight radiometric calibration experiment. Airborne remote sensing flights were carried out in northwestern Greenland in preparation for the Ice, Cloud, and land Elevation Satellite 2 (ICESat-2) laser altimeter mission. A total of nine science flights were conducted over the Greenland ice sheet, sea ice, and open-ocean water. The campaign's primary purpose was to correlate green laser pulse penetration into snow and ice with spectroscopic-derived surface properties. An experimental airborne instrument configuration that included a nadir-viewing (looking downward at the surface) non-imaging Analytical Spectral Devices (ASD) Inc. spectrometer that measured upwelling VSWIR (0.35 to 2.5 µ m) spectral radiance ( W m - 2 sr - 1 µ m - 1 "> W m − 2 sr − 1 µ m − 1 ) in the two-color Slope Imaging Multi-polarization Photon-Counting Lidar's (SIMPL) ground instantaneous field of view, and a zenith-viewing (looking upward at the sky) ASD spectrometer that measured VSWIR spectral irradiance (W m −2 nm −1 ) was flown. National Institute of Standards and Technology (NIST) traceable radiometric calibration procedures for laboratory, in-flight, and field environments are described in detail to achieve a targeted VSWIR measurement requirement of within 5 % to support calibration/validation efforts and remote sensing algorithm development. Our MODTRAN predictions for the 29 July flight line over dark and bright targets indicate that the airborne nadir-viewing spectrometer spectral radiance measurement uncertainty was between 0.6 % and 4.7 % for VSWIR wavelengths (0.4 to 2.0 µ m) with atmospheric transmittance greater than 80 %. MODTRAN predictions for Landsat 8 OLI relative spectral response functions suggest that OLI is measuring 6 % to 16 % more top-of-atmosphere (TOA) spectral radiance from the Greenland ice sheet surface than was predicted using apparent reflectance spectra from the nadir-viewing spectrometer. While more investigation is required to convert airborne VSWIR spectral radiance into atmospherically corrected airborne surface reflectance, it is expected that airborne science flight data products will contribute to spectroscopic determination of Greenland ice sheet surface optical properties to improve understanding of their potential influence on ICESat-2 measurements.

Atmospheric Measurement Techniques