Sources and magnitude of bias associated with determination of polychlorinated biphenyls in environmental samples

Abstract

Recently compiled data on the composition of commercial Aroclor mixtures and ECD (electron capture detector) response factors for all 209 PCB congeners are used to develop estimates of the bias associated with determination of polychlorinated blphenyis. During quantitation of multlcomponent peaks by congener-specific procedures error is introduced because of variable ECD response to isomeric PCBs. Under worst case conditions, the magnitude of this bias can range from less than 2% to as much as 600%. Multicomponent peaks containing the more highly and the lower chlorinated congeners experience the most bias. For this reason, quantitation of Σ PCB in Aroclor mixtures dominated by these species (e.g. 1016) are potentially subject to the greatest error. Comparison of response factor data for ECDs from two laboratories shows that the sign and magnitude of calibration bias for a given multicomponent peak is variable and depends, in part, on the response characteristics of individual detectors. By using the most abundant congener (of each multicomponent peak) for purposes of calibration, one can reduce the maximum bias to less than 55%. Moreover, due to cancellation of errors, the bias resulting from summation of all peak concentrations (i.e. Σ PCB) becomes vanishingly small (<1.2%). In contrast, bias associated with determination of Σ PCB as Aroclor equivalents by the traditional Aroclor method is potentially large (>200%) and highly variable in sign and magnitude. In this case, bias originates not only from the incomplete chromatographic resolution of PCB congeners but also the overlapping patterns of the Aroclor mixtures. Together these results illustrate the advantages of the congener-specific method of PCB quantitation over the traditional Aroclor Method and the extreme difficulty of estimating bias incurred by the latter procedure on a post hoc basis.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Robert P. Eganhouse, R. W. Gossett. 2002-05-01. Sources and magnitude of bias associated with determination of polychlorinated biphenyls in environmental samples. https://doi.org/10.1021/ac00019a012

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related USGS reports

Exploring the potential of electrospray-Orbitrap for stable isotope analysis using nitrate as a model

Widely used isotope ratio mass spectrometers have limited capabilities to measure metabolites, drugs, or small polyatomic ions without the loss of structural isotopic information. A new approach has recently been introduced that uses electrospray ionization Orbitrap to measure multidimensional isotope signatures of intact polar compounds. Using nitrate as a model compound, this study aims to establish performance metrics for comparisons with conventional IRMS at the natural abundance level. We present a framework on how to convert isotopolog intensities to δ values that are commonly used in the isotope geochemistry community. The quantification of seven nitrate isotopologs provides multiple pathways for obtaining the primary N and O δ values including non-mass-dependent O isotope variations, as well as opportunities to explore nonrandom isotopic distributions (i.e., clumping effects) within molecular nitrate. Using automation and the adaptation of measurement principles that are specific to isotope ratio analysis, nitrate δ15NAIR, δ18OVSMOW, and δ17OVSMOW were measured with a long-term precision of 0.4‰ or better for isotopic reference materials and purified nitrate from environmental samples. In addition, we demonstrate promising results for unpurified environmental samples in liquid form. With these new developments, this study connects the two largely disparate mass spectrometry fields of bioanalytical MS and isotope ratio MS, thus providing a route to measure new isotopic signatures in diverse organic and inorganic solutes.

Analytical Chemistry

New organic reference materials for hydrogen, carbon, and nitrogen stable isotope-ratio measurements: caffeines, n-alkanes, fatty acid methyl esters, glycines, L-valines, polyethylenes, and oils

An international project developed, quality-tested, and determined isotope−δ values of 19 new organic reference materials (RMs) for hydrogen, carbon, and nitrogen stable isotope-ratio measurements, in addition to analyzing pre-existing RMs NBS 22 (oil), IAEA-CH-7 (polyethylene foil), and IAEA-600 (caffeine). These new RMs enable users to normalize measurements of samples to isotope−δ scales. The RMs span a range of δ 2 H VSMOW-SLAP values from −210.8 to +397.0 mUr or ‰, for δ 13 C VPDB-LSVEC from −40.81 to +0.49 mUr and for δ 15 N Air from −5.21 to +61.53 mUr. Many of the new RMs are amenable to gas and liquid chromatography. The RMs include triads of isotopically contrasting caffeines, C 16 n -alkanes, n -C 20 -fatty acid methyl esters (FAMEs), glycines, and l -valines, together with polyethylene powder and string, one n -C 17 -FAME, a vacuum oil (NBS 22a) to replace NBS 22 oil, and a 2 H-enriched vacuum oil. A total of 11 laboratories from 7 countries used multiple analytical approaches and instrumentation for 2-point isotopic normalization against international primary measurement standards. The use of reference waters in silver tubes allowed direct normalization of δ 2 H values of organic materials against isotopic reference waters following the principle of identical treatment. Bayesian statistical analysis yielded the mean values reported here. New RMs are numbered from USGS61 through USGS78, in addition to NBS 22a. Because of exchangeable hydrogen, amino acid RMs currently are recommended only for carbon- and nitrogen-isotope measurements. Some amino acids contain 13 C and carbon-bound organic 2 H-enrichments at different molecular sites to provide RMs for potential site-specific isotopic analysis in future studies.

Analytical Chemistry

Statistical models for the analysis and design of digital polymerase chain (dPCR) experiments

Statistical methods for the analysis and design of experiments using digital PCR (dPCR) have received only limited attention and have been misused in many instances. To address this issue and to provide a more general approach to the analysis of dPCR data, we describe a class of statistical models for the analysis and design of experiments that require quantification of nucleic acids. These models are mathematically equivalent to generalized linear models of binomial responses that include a complementary, log&ndash;log link function and an offset that is dependent on the dPCR partition volume. These models are both versatile and easy to fit using conventional statistical software. Covariates can be used to specify different sources of variation in nucleic acid concentration, and a model&rsquo;s parameters can be used to quantify the effects of these covariates. For purposes of illustration, we analyzed dPCR data from different types of experiments, including serial dilution, evaluation of copy number variation, and quantification of gene expression. We also showed how these models can be used to help design dPCR experiments, as in selection of sample sizes needed to achieve desired levels of precision in estimates of nucleic acid concentration or to detect differences in concentration among treatments with prescribed levels of statistical power.

Analytical Chemistry