Geology ReportsSearch

USGS · 70010741

The detection of sulphur in contamination spots in electron probe X-ray microanalysis

Abstract

Sulphur has been identified as one of the elements present in the contamination spot which forms under the electron beam in the microprobe. The presence of the sulphur results in a rapid change in intensity measurements causing a loss of observed intensity for elements other than sulphur. The source of sulphur has been traced at least in part to the Apiezon B diffusion pump oil. A comparative X-ray fluorescence study of the Apiezon B and Octoil diffusion pump oils showed substantial amounts of sulphur in the Apiezon B. The Octoil was relatively free of sulphur.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

I. Adler, E.J. Dwornik, H. J. Rose. 2002-11-20. The detection of sulphur in contamination spots in electron probe X-ray microanalysis. https://doi.org/10.1088/0508-3443%2F13%2F5%2F314