Christopher S. Clayton

Publications and source records attributed to Christopher S. Clayton.

2 recordsLinked to original sources

Accuracy testing of steel and electric groundwater-level measuring tapes: Test method and in-service tape accuracy

The accuracy of groundwater-level tapes was investigated by developing a tape calibration method and device and testing the accuracy of a sample of groundwater-level tapes with the calibration method and device. The sample of tapes included in-service U.S. Geological Survey (USGS) Water Science Center steel and electric groundwater-level tapes. The tape calibration method developed during the study is based on a National Institute for Standards and Technology (NIST) protocol, and compares the tape under calibration (test tape) with a NIST-traceable reference tape. The calibration method can be used to determine tape accuracy and tape corrections. The tape calibration device consists of an anchor and tensioner. The device allows individual tensioning of the test and reference tapes. The Vernier scale on the tensioner allows measurement of the difference (or displacement) between the test and reference tapes graduations with a resolution of 0.0005 foot (ft). The calibration method used with the calibration device has a repeatability (standard deviation) of 0.0011 ft. The calibration device and proposed method were used to calibrate a sample of in-service USGS steel and electric groundwater tapes. The sample of in-service groundwater steel tapes were in relatively good condition. All steel tapes, except one, were accurate to ±0.01 ft per 100 ft over their entire length. One steel tape, which had obvious damage in the first hundred feet, was marginally outside the accuracy of ±0.01 ft per 100 ft by 0.001 ft. The sample of in-service groundwater-level electric tapes were in a range of conditions—from like new, with cosmetic damage, to nonfunctional. The in-service electric tapes did not meet the USGS accuracy recommendation of ±0.01 ft. In-service electric tapes, except for the nonfunctional tape, were accurate to about ±0.03 ft per 100 ft. A comparison of new with in-service electric tapes found that steel-core electric tapes maintained their length and accuracy better than electric tapes without a steel core. The in-service steel tapes could be used as is and achieve USGS accuracy recommendations for groundwater-level measurements. The in-service electric tapes require tape corrections to achieve USGS accuracy recommendations for groundwater-level measurement.

Open-File Report

Accuracy testing of electric groundwater-level measurement tapes

Electric tapes are used to measure groundwater levels and to verify the accuracy of pressure transducers installed in wells. Electric tapes are generally assumed to be accurate to ±0.01 foot (ft), but little information is available from the manufacturers and no accuracy studies have been conducted to confirm this value. This study measured the accuracy of six popular models of electric groundwater tapes. The tapes tested include models from Durham Geo, Geotech, Heron, In-Situ, Solinst, and Waterline that are commonly used by the U.S.Geological Survey (USGS). The accuracy tests compared the length of each electric tape to a calibrated-steel reference tape and measured each probe’s activation accuracy and displacement volume. The tape-length accuracy combined with the probe-activation accuracy gave the overall measurement accuracy of the tape. The accuracy tests demonstrated that none of the electric-tape models tested consistently met the suggested USGS accuracy of ±0.01 ft. The test data show that the tape models in the study should give a water-level measurement that is accurate to roughly ±0.05 ft per 100 ft without additional calibration. To meet USGS accuracy guidelines, the electric-tape models tested will need to be individually calibrated. Specific conductance also plays a part in tape accuracy. The probes will not work in water with specific conductance values near zero, and the accuracy of one probe was unreliable in very high conductivity water (10,000 microsiemens per centimeter).

Open-File Report